  • Instrument uses a 480x272 matrix 256 color LCD display, data storage 240 groups.
  • Advanced and reasonable opticalsystem design and 1,200lines / mm holographic gratings are guaranteed to achieve ultra-low stray light.
  • Advanced controlsystem for deuterium lamp, tungsten lamp life time monitoring.
  • Unique socket design deuterium and tungsten lamps, lamp replacement is free optics debugging.
  • The instrument can be automatic wavelength calibration, automatic wavelength setting, automatic switching light and dark current correction.
  • GLPself-identification function, testing instruments wavelength accuracy and photometric accuracy, and issue the testreports and power protection.

PDF Downloads

  • UV Visible Spectrophotometer
  • Espectrofotómetro Visible UV
  • Description

    Details

    Model AV-1600 AV-1800 AUV-1600 AUV-1800
    Wave Range 320-1100nm 190-1100nm
    Spectral Bandwidth 4nm 2nm 4nm 2nm
    Wavelength Accuracy ±0.5nm
    Wavelength Repeatability ≤0.2nm
    Photometric Accuracy ±0.002A(0~0.5A); ±0.004A(0.5~1A); ±0.5%T(0~100%T)
    Photometric Repeatability ≤0.001A(0~0.5A); ≤0.002A(0.5~1A); ≤0.2%T(0~100%T)
    Stray Light ≤0.05%T @ 360nm ≤0.05%T @ 220nm, 360nm
    Stability ±0.001A/h @ 500nm
    Baseline Flatness ±0.002A
    Noise ±0.0005A
    Display Way 480*272 large screen LCD
    Work Way T,A,C,E
    Wavelength Set Way Auto
    Display Range 0-200% T, -0.3-3A
    Detector Silicon photodiode
    Light Source deuterium and tungsten lamp
    Keyboard Film numeric button
    Data Output USB 2.0*2 USB1.0
    Print Output Parallel port
    Dimension
    (mm/inch)
    470x370x230 mm/19x15x9"
    Weight 14Kg/31LB